Highlights -
  • Flue gas analysis with up to 6 customizable gas sensors, changeable on-site by the user.
  • Extended measurement range for accurate readings even at high gas concentrations.
  • User-friendly, easy-to-change gas sensors and access to replaceable parts.
  • Sealed, impact-protected housing for durability in harsh environments.
  • Simple operation via Control Unit or easyEmission software (sold separately).
  • Removable operating and display unit, connectable via Bluetooth or data bus cable.

Technical Data -

Temperature - NTC  
   
Measuring range
-20 to +50 °C
Accuracy
±0.2 °C (-10 to +50 °C)
Resolution
0.1 °C (-20 to +50 °C)
Ambient temperature probe
   
Temperature - TC Type K (NiCr-Ni)  
   
Measuring range
-200 to +1370 °C
Accuracy
±1 °C (-200 to -100.1 °C)
±1 °C (200.1 to +1370 °C)
±0.4 °C (-100 to +200 °C)
Resolution
0.1 °C (-200 to +1370 °C)
 
   
Temperature - TC Type S (Pt10Rh-Pt)  
   
Measuring range
0 to +1760 °C
Accuracy
±1 °C (0 to +1760 °C)
Resolution
0.1 °C (0 to +1760 °C)
 
   
Differential Pressure - Piezoresistive  
   
Measuring range
-200 to +200 hPa
-40 to +40 hPa
Accuracy
±1.5 % of mv (+3 to +40 hPa)
±0.5 hPa (-49.9 to +49.9 hPa)
±1.5 % of mv (-200 to -50 hPa)
±0.03 hPa (-2.99 to +2.99 hPa)
±1.5 % of mv (-40 to -3 hPa)
±1.5 % of mv (+50 to +200 hPa)
Resolution
0.01 hPa (-40 to +40 hPa)
0.1 hPa (-200 to +200 hPa)
 
   
Absolute Pressure  
   
Measuring range
600 to +1150 hPa
Accuracy
±10 hPa
Resolution
1 hPa
 
   
Flue gas O₂  
   
Measuring range
0 to +25 Vol.%
Accuracy
±0.8% of fsv (0 to +25 Vol.%)
Resolution
0.01 Vol.% (0 to +25 Vol.%)
Reaction time t₉₀
20 s (t₉₅)
 
   
Flue gas CO (with H₂-compensation)  
   
Measuring range
0 to +10000 ppm
Accuracy
±10 ppm (0 to +199 ppm)
±5 % of mv (+200 to +2000 ppm)
±10 % of mv (+2001 to +10000 ppm)
Resolution
1 ppm (0 to +10000 ppm)
Reaction time t₉₀
40 s
H₂ only as an indicator
   
Flue gas COlow (with H₂-compensation)  
   
Measuring range
0 to 500 ppm
Accuracy
±5 % of mv (+40 to +500 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution
0.1 ppm (0 to +500 ppm)
Reaction time t₉₀
40 s
H₂ only as an indicator
   
Flue gas NO  
   
Measuring range
0 to +4000 ppm
Accuracy
±5 % of mv (+100 to +1999 ppm)
±10 % of mv (+2000 to +4000 ppm)
±5 ppm (0 to +99 ppm)
Resolution
1 ppm (0 to +4000 ppm)
Reaction time t₉₀
30 s
 
   
Flue gas Nolow  
   
Measuring range
0 to +300 ppm
Accuracy
±5 % of mv (+40 to +300 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution
±0.1 ppm (0 to +300 ppm)
Reaction time t₉₀
30 s
 
   
Flue gas NO₂  
   
Measuring range
0 to +500 ppm
Accuracy
±5 % of mv (+100 to +500 ppm)
±5 ppm (0 to +99.9 ppm)
Resolution
0.1 ppm (0 to +500 ppm)
Reaction time t₉₀
40 s
 
   
Flue gas SO₂  
   
Measuring range
0 to +5000 ppm
Accuracy
±10 % of mv (+2001 to +5000 ppm)
±5 % of mv (+100 to +2000 ppm)
±5 ppm (0 to +99 ppm)
Resolution
1 ppm (0 to +5000 ppm)
Reaction time t₉₀
30 s
 
   
Flue gas degree of effectivity, Eta (calculated)  
   
Measuring range
0 to +120 %
Resolution
0.1 % (0 to +120 %)
 
   
Flue gas loss (calculated)  
   
Measuring range
0 to +99.9 %
Resolution
0.1 % (0 to +99.9 %)
 
   
Flue gas dewpoint (calculated)  
   
Measuring range
0 to +99.9 °Ctd
Resolution
0.1 °Ctd (0 to +99.9 °Ctd)
 
   
Flue gas CO₂ calculation (calculated from O₂)  
   
Measuring range
0 to CO₂ max
Accuracy
calculated from O₂ ±0.2 Vol.%
Resolution
0.01 Vol.%
 
   
Flue gas H₂S  
   
Measuring range
0 to +300 ppm
Accuracy
±5 % of mv (+40 to +300 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution
0.1 ppm (0 to +300 ppm)
Reaction time t₉₀
35 s
 
   
Flue gas CO₂ - NDIR  
   
Measuring range
0 to +50 Vol.%
Accuracy
±0.5 Vol.% + 1.5 % of mv (>25 to 50 Vol.%)
±0.3 Vol.% + 1 % of mv (0 to 25 Vol.%)
Resolution
0.01 Vol.% (0 to 25 Vol.%)
0.1 Vol.% (> 25 Vol.%)
Reaction time t₉₀
10 s
 
   
Velocity / Volume flow  
   
Measuring range
0 to +40 m/s
Resolution
0.1 m/s (0 to +40 m/s)
 
   
Methane (CH₄)  
   
Measuring range
100 to 40000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (2 x mV methane)
Reaction time t₉₀
< 40 s
Response factor
1
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Propane (C₃H₈)  
   
Measuring range
100 to 21000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (5 x mV propane)
Reaction time t₉₀
< 40 s
Response factor
1.5
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Butane (C₄H₁₀)  
   
Measuring range
100 to 18000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (6.5 x mV butane)
Reaction time t₉₀
< 40 s
Response factor
2
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Flue gas SO₂ low  
   
Measuring range
0 to 200 ppm
Accuracy
±5 (40 to 200 ppm)
±2 ppm (0 to 39.9 ppm)
Resolution
0.1 ppm
Reaction time t₉₀
30 s
In combination with the special SO2low gas sampling probe 0563 1251 or the industrial heated probe kit 0600 7630
   
General technical data  
   
Weight
4800 g
Dimensions
330 x 128 x 438 mm
Operating temperature
-5 to +45 °C
Product-/housing material
ABS
Protection class
IP40
Connectable probes
1 x combustion air temperature probe; 1 x flue gas probe; 1 x temperature probe; 1 x differential pressure
Product colour
Black
Battery life
5 h
Battery type
lithium batteries
DC electric tension input
11V to 40V
Power supply
AC mains unit 100V to 240V (50 to 60 Hz)
Dewpoint calculation
0 °Ctd; 99 °Ctd
Hose length
16.2 m (corresponds to 5 probe hose extensions)
Max. dust load
20 g/m³ dust in flue gas
Max. humidity load
70 °C dewpoint temperature
Max. neg. press./flue gas
-300 mbar
Max. pos. press./flue gas
50 mbar
Pump flow
1 l/min. with through-flow monitoring
Trigger input
Voltage 5 to 12 Volt (rising or trailing edge) pulse width > 1 sec load: 5 V/max, 5 mA, 12 V/max. 40 mA
Maximum memory
250,000 readings
Interface
data bus; Bluetooth®; USB; IR/IRDA interface; Mains connection; trigger input; Differential Pressure; probe input/ probe inputs; DC-voltage input; dilution gas inlet for measurement range extension
Storage temperature
-20 to +50 °C


View full details
Highlights -
  • Flue gas analysis with up to 6 customizable gas sensors, changeable on-site by the user.
  • Extended measurement range for accurate readings even at high gas concentrations.
  • User-friendly, easy-to-change gas sensors and access to replaceable parts.
  • Sealed, impact-protected housing for durability in harsh environments.
  • Simple operation via Control Unit or easyEmission software (sold separately).
  • Removable operating and display unit, connectable via Bluetooth or data bus cable.

Technical Data -
Temperature - NTC  
   
Measuring range -20 to +50 °C
Accuracy ±0.2 °C (-10 to +50 °C)
Resolution 0.1 °C (-20 to +50 °C)
Ambient temperature probe
   
Temperature - TC Type K (NiCr-Ni)  
   
Measuring range -200 to +1370 °C
Accuracy ±1 °C (-200 to -100.1 °C)
±1 °C (200.1 to +1370 °C)
±0.4 °C (-100 to +200 °C)
Resolution 0.1 °C (-200 to +1370 °C)
 
   
Temperature - TC Type S (Pt10Rh-Pt)  
   
Measuring range 0 to +1760 °C
Accuracy ±1 °C (0 to +1760 °C)
Resolution 0.1 °C (0 to +1760 °C)
 
   
Differential Pressure - Piezoresistive  
   
Measuring range -200 to +200 hPa
-40 to +40 hPa
Accuracy ±1.5 % of mv (+3 to +40 hPa)
±0.5 hPa (-49.9 to +49.9 hPa)
±1.5 % of mv (-200 to -50 hPa)
±0.03 hPa (-2.99 to +2.99 hPa)
±1.5 % of mv (-40 to -3 hPa)
±1.5 % of mv (+50 to +200 hPa)
Resolution 0.01 hPa (-40 to +40 hPa)
0.1 hPa (-200 to +200 hPa)
 
   
Absolute Pressure  
   
Measuring range 600 to +1150 hPa
Accuracy ±10 hPa
Resolution 1 hPa
 
   
Flue gas O₂  
   
Measuring range 0 to +25 Vol.%
Accuracy ±0.8% of fsv (0 to +25 Vol.%)
Resolution 0.01 Vol.% (0 to +25 Vol.%)
Reaction time t₉₀ 20 s (t₉₅)
 
   
Flue gas CO (with H₂-compensation)  
   
Measuring range 0 to +10000 ppm
Accuracy ±10 ppm (0 to +199 ppm)
±5 % of mv (+200 to +2000 ppm)
±10 % of mv (+2001 to +10000 ppm)
Resolution 1 ppm (0 to +10000 ppm)
Reaction time t₉₀ 40 s
H₂ only as an indicator
   
Flue gas COlow (with H₂-compensation)  
   
Measuring range 0 to 500 ppm
Accuracy ±5 % of mv (+40 to +500 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution 0.1 ppm (0 to +500 ppm)
Reaction time t₉₀ 40 s
H₂ only as an indicator
   
Flue gas NO  
   
Measuring range 0 to +4000 ppm
Accuracy ±5 % of mv (+100 to +1999 ppm)
±10 % of mv (+2000 to +4000 ppm)
±5 ppm (0 to +99 ppm)
Resolution 1 ppm (0 to +4000 ppm)
Reaction time t₉₀ 30 s
 
   
Flue gas Nolow  
   
Measuring range 0 to +300 ppm
Accuracy ±5 % of mv (+40 to +300 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution ±0.1 ppm (0 to +300 ppm)
Reaction time t₉₀ 30 s
 
   
Flue gas NO₂  
   
Measuring range 0 to +500 ppm
Accuracy ±5 % of mv (+100 to +500 ppm)
±5 ppm (0 to +99.9 ppm)
Resolution 0.1 ppm (0 to +500 ppm)
Reaction time t₉₀ 40 s
 
   
Flue gas SO₂  
   
Measuring range 0 to +5000 ppm
Accuracy ±10 % of mv (+2001 to +5000 ppm)
±5 % of mv (+100 to +2000 ppm)
±5 ppm (0 to +99 ppm)
Resolution 1 ppm (0 to +5000 ppm)
Reaction time t₉₀ 30 s
 
   
Flue gas degree of effectivity, Eta (calculated)  
   
Measuring range 0 to +120 %
Resolution 0.1 % (0 to +120 %)
 
   
Flue gas loss (calculated)  
   
Measuring range 0 to +99.9 %
Resolution 0.1 % (0 to +99.9 %)
 
   
Flue gas dewpoint (calculated)  
   
Measuring range 0 to +99.9 °Ctd
Resolution 0.1 °Ctd (0 to +99.9 °Ctd)
 
   
Flue gas CO₂ calculation (calculated from O₂)  
   
Measuring range 0 to CO₂ max
Accuracy calculated from O₂ ±0.2 Vol.%
Resolution 0.01 Vol.%
 
   
Flue gas H₂S  
   
Measuring range 0 to +300 ppm
Accuracy ±5 % of mv (+40 to +300 ppm)
±2 ppm (0 to +39.9 ppm)
Resolution 0.1 ppm (0 to +300 ppm)
Reaction time t₉₀ 35 s
 
   
Flue gas CO₂ - NDIR  
   
Measuring range 0 to +50 Vol.%
Accuracy ±0.5 Vol.% + 1.5 % of mv (>25 to 50 Vol.%)
±0.3 Vol.% + 1 % of mv (0 to 25 Vol.%)
Resolution 0.01 Vol.% (0 to 25 Vol.%)
0.1 Vol.% (> 25 Vol.%)
Reaction time t₉₀ 10 s
 
   
Velocity / Volume flow  
   
Measuring range 0 to +40 m/s
Resolution 0.1 m/s (0 to +40 m/s)
 
   
Methane (CH₄)  
   
Measuring range 100 to 40000 ppm
Accuracy < 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution 10 ppm
Min. O₂ requirement in flue gas 2 % + (2 x mV methane)
Reaction time t₉₀ < 40 s
Response factor 1
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Propane (C₃H₈)  
   
Measuring range 100 to 21000 ppm
Accuracy < 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution 10 ppm
Min. O₂ requirement in flue gas 2 % + (5 x mV propane)
Reaction time t₉₀ < 40 s
Response factor 1.5
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Butane (C₄H₁₀)  
   
Measuring range 100 to 18000 ppm
Accuracy < 400 ppm (100 to 4000 ppm)
< 10 % of mv (> 4000 ppm)
Resolution 10 ppm
Min. O₂ requirement in flue gas 2 % + (6.5 x mV butane)
Reaction time t₉₀ < 40 s
Response factor 2
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
   
Flue gas SO₂ low  
   
Measuring range 0 to 200 ppm
Accuracy ±5 (40 to 200 ppm)
±2 ppm (0 to 39.9 ppm)
Resolution 0.1 ppm
Reaction time t₉₀ 30 s
In combination with the special SO2low gas sampling probe 0563 1251 or the industrial heated probe kit 0600 7630
   
General technical data  
   
Weight 4800 g
Dimensions 330 x 128 x 438 mm
Operating temperature -5 to +45 °C
Product-/housing material ABS
Protection class IP40
Connectable probes 1 x combustion air temperature probe; 1 x flue gas probe; 1 x temperature probe; 1 x differential pressure
Product colour Black
Battery life 5 h
Battery type lithium batteries
DC electric tension input 11V to 40V
Power supply AC mains unit 100V to 240V (50 to 60 Hz)
Dewpoint calculation 0 °Ctd; 99 °Ctd
Hose length 16.2 m (corresponds to 5 probe hose extensions)
Max. dust load 20 g/m³ dust in flue gas
Max. humidity load 70 °C dewpoint temperature
Max. neg. press./flue gas -300 mbar
Max. pos. press./flue gas 50 mbar
Pump flow 1 l/min. with through-flow monitoring
Trigger input Voltage 5 to 12 Volt (rising or trailing edge) pulse width > 1 sec load: 5 V/max, 5 mA, 12 V/max. 40 mA
Maximum memory 250,000 readings
Interface data bus; Bluetooth®; USB; IR/IRDA interface; Mains connection; trigger input; Differential Pressure; probe input/ probe inputs; DC-voltage input; dilution gas inlet for measurement range extension
Storage temperature -20 to +50 °C